.pdf File
CD Start Page DATA ACQUISITION ONLINE Page DATA ACQUISITION PRODUCTS Page

 

OM-3000 From $ 995
Temperature or Voltage Portable Datalogging System

OM-3000

Description
The OM-300 datalogger will accept direct input from up to 6 thermocouples or 4 voltage inputs, which can be programmed through the integral keypad. The inputs may be mixed, to a maximum of 6 inputs, provided that only up to 4 of them are voltage. For instance, the OM-3000 can be programmed for 6 thermocouple inputs, or for 4 voltage inputs and 2 thermocouple inputs. The OM-3000 will accept thermocouple types, E, J, K, and T, as well as voltage ranges of +5 Vdc of +30 Vdc, with a common mode voltage range of +30 Vdc. A large 4.3" x 2.3" integral backlit display with graphic capabilities can be used for on-the-spot measurements for all 6 inputs simultaneously. Input ranges, user-defined session name and sampling intervals are easily programmed via the built-in keypad.

Taking measurements and downloading them to your PC for further analysis or archiving has never been easier. The OM-3000 comes standard with measurement and analysis software for IBM or compatible PC's which is a sophisticated yet extremely user-friendly program which can be used to measure inputs on all 6 channels, including min and max, bar charts and Y vs time graphs. Sampling intervals may be programmed from 0.2 seconds to 60 hours (for all 6 channels at once). Approximately 100,000 individuals samples may best stored in the OM-3000 for downloading. Each recording has a unique time/date stamp, and a user specified name can be entered from the integral keypad.

Communications is established through an RS-232 serial port on the OM-3000, which may be programmed for baud rates from 9600 to 57,600. A serial cable with a 25 pin female DB-25 connector is included with each OM-3000. If connecting to a 9-pin serial port on an IBM or compatible PC, a 9 to 25 pin adapter is required.

The OM-3000 is powered from a rechargeable NiCAD battery pack within 800mA hour rating at 7.2 Vdc. The OM-3000 will operate from up to 8 hours, with backlight off and up to 5 hours with backlight on with a full charged battery. The unit may operate continuously if left connected to AC power. An integral dual rate charger is capable of charging the NiCad battery pack in approximately 1.25 hours. In addition, the internal battery combats NiCad memory effect.

PRICING
Model No. Price Description
OM-3000 $995 Portable thermocouple and voltage datalogger
OM-3000-SC 15 Soft fabric carrying case. Will contain logger and accessories

Each unit is supplied with Windows software, battery charger plug pack,
RS-232 cable and complete operator's manual.
Ordering Example: OM-3000, portable thermocouple and voltage datalogger, plus
OM-3000-SC soft fabric carrying case, $995- + 15 = $1010.


COPY [To Your Hard Drive] OM-3000 Windows software now!

VIEW an Adobe Acrobat pdf file for more product info (requires Acrobat Reader).


Sales and Service 1.800.826.6342
U.S.A. and Canada
International Callers;
Sales and Service; 203.359.1660
Please E-Mail any comments or suggestions to OMEGA®.
www.omega.com
® COPYRIGHT 1999 OMEGA ENGINEERING, INC.